• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM; CORONARY ATHERECTOMY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM; CORONARY ATHERECTOMY DEVICE Back to Search Results
Model Number DBEC-125
Device Problems Unintended System Motion (1430); Unintended Movement (3026)
Patient Problems Low Blood Pressure/ Hypotension (1914); Ischemia (1942)
Event Date 08/29/2022
Event Type  Injury  
Manufacturer Narrative
The device history record for the reported oad could not be reviewed as the lot number was not provided.If the lot number is provided, a dhr review will be performed.The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.Csi id: (b)(4).
 
Event Description
The crown became stuck and the diamondback 360 coronary orbital atherectomy device (oad) stalled in the left main coronary artery (lm), which resulted in hypotension and severe ischemia.The device was forcefully extracted, and wire position was lost.The physician then advanced a low-profile balloon, followed by intravascular lithotripsy (ivl) and stent placement.The patient was stabilized with vasopressors and fluids.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM
Type of Device
CORONARY ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
aaron stevens
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key15502305
MDR Text Key300759839
Report Number3004742232-2022-00230
Device Sequence Number1
Product Code MCX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDBEC-125
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-