• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. DISCOFIX®; ACCESSORIES, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. DISCOFIX®; ACCESSORIES, CATHETER Back to Search Results
Model Number 456020
Device Problem Leak/Splash (1354)
Patient Problems Awareness during Anaesthesia (1707); Low Oxygen Saturation (2477)
Event Date 08/26/2022
Event Type  malfunction  
Event Description
As reported by the user facility: stopcock was leaking and medications for sedation spilled onto bed.Nurse noticed pt to be interacting and the ventilator unexpectedly a sudden (down arrow) in spo2 and (up arrow) in agitation despite (changes) being made to pt's sedation.When nurse looked at pt's iv line, she realized that the stopcock was leaking propofol and fentanyl and pt's sheets soaked and medications that is supposed to be insuing in the iv line.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).One (1) unused sample with packaging was returned for evaluation.Sample was subject to a visual evaluation with passing results.The sample was tested as per specification and no leakage was observed.Based on the evaluation results, the reported defect was not confirmed.Retained units were evaluated and passed the internal testing.A review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.Additionally, review of the batch history records was also performed for the reported lot number and no non-conformances or deviations were noted during the manufacturing process or final inspections.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISCOFIX®
Type of Device
ACCESSORIES, CATHETER
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key15503595
MDR Text Key300788267
Report Number2523676-2022-00484
Device Sequence Number1
Product Code KGZ
UDI-Device Identifier04046964996962
UDI-Public(01)04046964996962
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K760383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number456020
Device Catalogue Number456020
Device Lot Number21G09D9045
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/15/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/10/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-