MEDOS INTERNATIONAL SARL TRUESPAN MENISCAL REPAIR SYSTEM PLGA 24 DEGREE; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Model Number 228162
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2022
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).Reporter is a j&j sales representative.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event Description
It was reported by the sales rep in switzerland that during a meniscal repair procedure on (b)(6) 2022, it was observed that there was a knot on the suture on the truespan meniscal repair system plga 24 degree device.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was nt available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the complaint device was received and evaluated.Upon visual inspection, only a piece of 3.5 cm.Was returned.A knot on the suture can be observed.The complete device was not returned.A manufacturing record evaluation was performed for the finished device 9l35934 number, and no non-conformances were identified.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection, this complaint can be confirmed.Multiple factors can be associated with this type of failure, the customer did not provide further information in regards the issue.The complete device was not returned, therefore we cannot perform a full evaluation of the device and provide a root cause of the failure.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUESPAN MENISCAL REPAIR SYSTEM PLGA 24 DEGREE
Type of Device
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15504079
MDR Text Key306254526
Report Number1221934-2022-02974
Device Sequence Number1
Product Code MBI
UDI-Device Identifier10886705026050
UDI-Public10886705026050
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K153667
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number228162
Device Catalogue Number228162
Device Lot Number9L35934
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/21/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-