MEDOS INTERNATIONAL SARL TRUESPAN MENISCAL REPAIR SYSTEM PEEK 24 DEGREE; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Model Number 228152
Device Problems Device-Device Incompatibility (2919); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/19/2022
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event Description
This is report 2 of 2 for (b)(4).It was reported by a healthcare professional in china that during a meniscal repair procedure on (b)(6) 2022, it was observed that the two plates on the truespan meniscal repair system peek 24 degree device were deployed at the same time and the needle was deformed after the first firing.They changed to another one but the same issue occurred.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary : the product was returned to mitek for evaluation.Mitek then conducted visual inspection and functional test of device received.The visual inspection of the second device revealed that both implants were deployed, and they were still in the suture.The suture was not in the needle.The red trigger was reviewed and tested, it performed as intended.The pusher rod that places the implants is also in good shape, no structural anomalies could be found.The device has no structural damages on its body.Manufacturing record evaluation was performed for the finished device lot number: 9l11309, and no nonconformances were identified.According with the visual inspection of the implants, this complaint can be confirmed.The possible root cause for the deployment failure reported could be related to a trigger mishandling, if the trigger was activated unintended, the deployment mechanism starts its process and when the trigger was used in the procedure, both implants would be deployed at the same time in the first shoot.As per ifu, it is necessary to use a calibrated probe, measure the width of the meniscal tissue to be repaired and set the adjustable depth.Also, during needle insertion use a malleable graft retractor or slotted cannula to prevent the needle from catching on or damaging soft tissue; once inside the joint space, remove the instrument.At this point in time, no corrective action is required, and no further action is warranted.However, in depuy synthes mitek, additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as apropriate.H10 additional narrative: investigation summary:photos were received and evaluated.Upon visual inspection of the photos, it appeared that the needle of one of the devices was deformed, also the other photo shows both implants deployed outside of the gun; therefore, this complaint can be confirmed for all the reported issues.A manufacturing record evaluation was performed for the finished device lot number: 9l11309, and no nonconformances were identified.The photos do not provide enough evidence to determine root cause.Without physical evaluation of the devices reported, we cannot establish a root cause for the issue experienced by the customer.The possible root cause for the bent needle can be attributed to procedural variables, such handling of the device or product interaction during procedure; the needle may have been forced upwards while it was inserted into the soft tissue.The possible root cause for the deployment failure could be related when not inserting the needle to the proper depth for deployment which have caused blocking insertion of the first implant, and when this occurred may have felt resistance and did not pull the trigger all the way.This would result in the first implant being only partially deployed.Then, when the trigger was pulled again both implants would be deployed at the same time, however, this cannot be conclusively determined.As per ifu, it is necessary to use a calibrated probe, measure the width of the meniscal tissue to be repaired and set the adjustable depth.At desired depth, fully squeeze the red deployment trigger while maintaining depth positioning to deliver the first implant.The implant is fully deployed when you hear an audible ¿click¿.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUESPAN MENISCAL REPAIR SYSTEM PEEK 24 DEGREE
Type of Device
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15509362
MDR Text Key300870000
Report Number1221934-2022-02995
Device Sequence Number1
Product Code MBI
UDI-Device Identifier10886705026029
UDI-Public10886705026029
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K153667
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number228152
Device Catalogue Number228152
Device Lot Number9L11309
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/22/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-