Reportable based on device analysis completed on (b)(6) 2022.It was reported that device damage occurred.A 32 x 2.50 rebel bms stent was selected for used.However, the hospital informed that the device is defective.There were no patient complications reported.However, returned device analysis revealed shaft detached.
|
Date of event: used the first date of the month of the aware date as no event date was provided.Returned product consisted of the rebel bms stent delivery system with stent.The shafts, tip, stent, and balloon were microscopically and visually examined.There was contract in the inflation lumen and blood in the guidewire lumen.The balloon was tightly folded with the stent attached and secure between the markerbands.Inspection of the device revealed that there were numerous kinks throughout the hypotube and the hypotube was separated 32cm distal of the strain relief.The ends of the separation were ovaled, indicating that the hypotube was kinked prior to separation.
|