The company service representative, examined the system.And was unable to confirm or replicate the reported event.The system software was upgraded, and systems checks were performed.The system was then tested and met all product specifications.A system non-conformance based review of the batch/lot/serial number was performed and did not reveal, any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed, prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this lot/batch/serial number was performed.No similar complaints were reported for the product lot/batch/serial under investigation.The system was found to functionally exhibit no problems.Therefore, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|