• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION REZUM; UNIT, ELECTROSURGICAL ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION REZUM; UNIT, ELECTROSURGICAL ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number D2201
Device Problem Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/12/2022
Event Type  malfunction  
Event Description
It was reported that, following the priming cycle, the saline continued to flow although the needle was retracted.It was observed that the wrong scope was inserted.As the scope was removed, it was noted that the shaft disconnected from the handpiece but the silver pin was still intact.A second delivery device was used to complete the procedure without clinical consequences to the patient.
 
Event Description
It was reported that, following the priming cycle, the saline continued to flow although the needle was retracted.It was observed that the length of the scope was longer than the delivery device, the scope that was attempted to be used was the wrong scope for the delivery device.As the scope was removed, it was noted that the shaft disconnected from the handpiece but the silver pin was still intact.A second delivery device was used to complete the procedure without clinical consequences to the patient.
 
Manufacturer Narrative
Based on the available information, the root cause of the reported clinical observation cannot be established, as the product has not been returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REZUM
Type of Device
UNIT, ELECTROSURGICAL ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
alyson harris
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key15519848
MDR Text Key304135168
Report Number2124215-2022-38092
Device Sequence Number1
Product Code KNS
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberD2201
Device Catalogue NumberD2201
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-