• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH AVALON FM30 FETAL MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH AVALON FM30 FETAL MONITOR Back to Search Results
Model Number M2703A
Device Problems Circuit Failure (1089); Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/23/2022
Event Type  malfunction  
Event Description
It was reported that there was no sound although an error appear on the screen.The device was in use on a patient.There was no report of patient or user harm.
 
Manufacturer Narrative
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
 
Event Description
The customer reported an avalon fm30 was displaying an error message and had no sound.The device was in use on a patient.There was no report of patient or user harm.The remote service engineer (rse) performed troubleshooting with the customer and confirmed the issue was a bus master malfunction.A philips field service engineer (fse) was sent onsite to replace the bus master board to resolve the issue.After the board was replaced the error message went away.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVALON FM30 FETAL MONITOR
Type of Device
AVALON FM30 FETAL MONITOR
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
jeanne ahearn
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key15539662
MDR Text Key301134331
Report Number9610816-2022-00509
Device Sequence Number1
Product Code HGM
UDI-Device Identifier00884838000414
UDI-Public00884838000414
Combination Product (y/n)N
Reporter Country CodeSA
PMA/PMN Number
K140535
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberM2703A
Device Catalogue NumberM2703A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/19/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/17/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-