Model Number N/A |
Device Problem
Packaging Problem (3007)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/19/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that there was a hair on an articular surface in the sterile packaging.The implant was not used.Attempts to obtain additional information have been made; however, no more information is available.
|
|
Manufacturer Narrative
|
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation as it was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Manufacturer Narrative
|
This follow-up is being submitted to relay additional information.Visual evaluation of the provided photo found a hair-like debris on the device.The sterile packaging was opened prior to the discovery of the debris, and therefore, the source of the debris cannot be confirmed.Medical records were not provided.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.A definitive root cause cannot be determined.If any further information is which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Search Alerts/Recalls
|