• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TIBIAL COMPONENT PRECOAT RIGHT MEDIAL/LEFT LATERAL SIZE 3; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TIBIAL COMPONENT PRECOAT RIGHT MEDIAL/LEFT LATERAL SIZE 3; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problems Unstable (1667); Noise, Audible (3273); Loosening of Implant Not Related to Bone-Ingrowth (4002); Migration (4003)
Patient Problems Failure of Implant (1924); Pain (1994); Osteolysis (2377); Ambulation Difficulties (2544); Swelling/ Edema (4577)
Event Date 08/29/2022
Event Type  Injury  
Manufacturer Narrative
(b)(4).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2022-02890.
 
Event Description
It was reported a patient underwent a right unicompartmental knee arthroplasty.Approximately 14 years post implantation, the patient underwent a conversion to a total knee arthroplasty due to pain, instability, swelling, difficulty with ambulation, and radiographic evidence of subsidence.During the revision, aseptic loosening of the tibial component was found.All implants were removed and replaced.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.The device history record was unable to be performed as the lot number of the device involved in the event is unknown.Medical records/radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: patient have experienced pain, clicking, instability, swelling with difficult ambulation.A revision surgery was performed as loosening and subsidence was noted.X-ray was reviewed and noted loosening and subsidence of the tibial implant with extensive tibial osteolysis and knee varus alignment.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL COMPONENT PRECOAT RIGHT MEDIAL/LEFT LATERAL SIZE 3
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15544930
MDR Text Key301202939
Report Number0001822565-2022-02889
Device Sequence Number1
Product Code HSX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K033363
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/05/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00584200302
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexMale
Patient Weight76 KG
-
-