Model Number 72204041
Device Problems Break (1069); Failure to Fire (2610)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/07/2022
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that during an acl reconstruction procedure, the tip of blade in the 7.5mm retrograde drill could not deployed and the wire that retracts the cutter blade was broken when the 4.9mm bone hole was made; after retrograde drilling/before cutter retraction the broken pieces removed from patient.The procedure was successfully completed with non-significant surgical delay using a back-up device.No further complications were reported.
 
Manufacturer Narrative
H10: internal complaint reference: (b)(4).H3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A visual inspection of the returned device found that it is not in its original packaging.The retrograde bit is fractured from it's window within the antegrade bit and completely unattached.The actuator wire is deformed at the distal end and unattached from the grey slider at the proximal end.There is debris throughout the returned items.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force during use, failure to retract the guidewire far enough back before attempting to deploy the blade, using the reverse function while drilling, attempting to deploy the blade while still inside the bone tunnel, or an impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RETROGRADE DRL 7.5MM
Type of Device
BIT, SURGICAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15545352
MDR Text Key301206256
Report Number1219602-2022-01482
Device Sequence Number1
Product Code GFG
UDI-Device Identifier00885554037708
UDI-Public00885554037708
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/05/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72204041
Device Catalogue Number72204041
Device Lot Number4954642
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/17/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/25/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-