• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ENTERRA; INTESTINAL STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ENTERRA; INTESTINAL STIMULATOR Back to Search Results
Model Number 37800
Device Problems Entrapment of Device (1212); Migration or Expulsion of Device (1395)
Patient Problems Erosion (1750); Foreign Body In Patient (2687)
Event Date 08/11/2021
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: product id 4351-35 lot# (b)(4) serial# implanted: (b)(6) 2018 explanted: (b)(6) 2021 product type lead product id 4351-35 lot# (b)(4) serial# implanted: (b)(6) 2018 explanted: (b)(6) 2021 product type lead.Other relevant device(s) are: product id: 4351-35, serial/lot #: (b)(4), ubd: 27-sep-2019, udi#: (b)(4); product id: 4351-35, serial/lot #: (b)(4), ubd: 28-jul-2019, udi#: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare provider regarding a patient who was implanted with an implantable neurostimulator (ins) for gastric stimulation.It was reported that the patients device was replaced due to eroded gastric stimulator leads. they spent about 2 hours lysing adhesions around the anterior abdominal wall.Had a lot of fibrotic reaction around the leads.The leads seem to go around a few loops of small bowel.The leads were found and removed successfully.On (b)(6) 2021, at their follow up appointment, the patient was not wearing the binder and there was some fluid around the ins.Doctor advised patient to wear binder to prevent infection.There was some mild serous drainage from the umbilicus, so the doctor opened this with silver nitrate in order to prevent an infection from developing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key15555216
MDR Text Key301334060
Report Number3004209178-2022-12993
Device Sequence Number1
Product Code LNQ
UDI-Device Identifier00763000286798
UDI-Public00763000286798
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H99001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 10/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/14/2021
Device Model Number37800
Device Catalogue Number37800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/30/2022
Date Device Manufactured06/23/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Life Threatening;
Patient Age62 YR
Patient SexFemale
-
-