MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. OES CYSTONEPHROFIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. OES CYSTONEPHROFIBERSCOPE Back to Search Results
Model Number CYF-5
Device Problem Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/06/2022
Event Type  malfunction  
Event Description
The customer reported the working channel port came off of the cystonephrofiberscope.The customer noted the scope had just been returned from repair and the issue was found during cleaning.There was no procedure at the time of this event.There was no reported patient harm or impact due to this event.
 
Manufacturer Narrative
During device evaluation at olympus, it was found biopsy port was missing and there was leaks coming from a hole in the bending section cover and instrument channel.The battery mark was also missing.This event is under investigation.A supplemental report will be submitted upon receiving additional information.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 9 years since the subject device was manufactured.Based on the results of the investigation, the detachment of the forceps port was likely caused by the same cause as other damages, but it could not be determined whether the detachment was due to stress, handling, or other factors.A definitive root cause cannot be identified.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES CYSTONEPHROFIBERSCOPE
Type of Device
CYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15555443
MDR Text Key306907238
Report Number9610595-2022-02685
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170339417
UDI-Public04953170339417
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K032092
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2022
Was the Report Sent to FDA? No
Date Manufacturer Received11/01/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/04/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-