DEPUY ORTHOPAEDICS INC US PINN POLY IMPACTOR TIP 36MM; HIP INSTRUMENTS : INSERTION DEVICES Back to Search Results
Model Number 2217-50-008
Device Problems Crack (1135); Device-Device Incompatibility (2919); Naturally Worn (2988)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/23/2022
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the impactors were worn and cracking.No surgical delay was reported.
 
Manufacturer Narrative
Product complaint #
=
> pc-(b)(4).Investigation summary
=
> the device associated with this report was received for examination.Visual inspection of the returned device found the device overall appearance is worn.The reported worn condition was confirmed.However, no signs of cracks were found.A functional test performed with mating sample was unable to replicate the unable to assemble condition.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot
=
> a date code was provided, which indicates that the device was manufactured on (5/2005).A manufacturing records evaluation (mre) was not performed since a valid finished good lot number was not provided for this device.Corrected: h3.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Additional information received indicated that the instrument was showing signs of crack on the colored impaction part/ unthreading from silver to attach to handle.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN POLY IMPACTOR TIP 36MM
Type of Device
HIP INSTRUMENTS : INSERTION DEVICES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15561151
MDR Text Key306507432
Report Number1818910-2022-19751
Device Sequence Number1
Product Code HXC
UDI-Device Identifier10603295098973
UDI-Public10603295098973
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 10/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2217-50-008
Device Catalogue Number221750008
Device Lot NumberAY0505
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/15/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-