Model Number 187954 |
Device Problems
Material Disintegration (1177); Difficult to Remove (1528)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that dressing was much more difficult to remove from skin when changing dressings than it used to be.The dressing got stuck extremely strongly, sometimes only be removed in pieces for which more use of adhesive remover required significantly.Along with the company's dressing, the treatment regimen of the practitioners provided for the additional use of a foil bandage to minimize the rolling up of the bandage edges of dressing so far with complete satisfaction.Although the products were heavier and more time-consuming to use, the customer continued to use the products before and after reporting the detected quality change as they saw no risk to the patients.According to the customer, he reported this issue for several uses and due to the accumulation of difficult handling of the product.The difficulty in detaching from the skin was independent of the skin type.The product was used for two to four days.There was no harm reported.The photographs depicting the issues were received from the complainant.
|
|
Manufacturer Narrative
|
Device 3 of 22.Common device name ¿ dressing, wound, occlusive.Complainant city: (b)(6).Complainant state: (b)(6).Complainant postal code: (b)(6).Complainant country: (b)(6).Name of affiliation: (b)(6).(b)(4).Based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.(b)(4).
|
|
Event Description
|
To date no additional patient or event details have been received.
|
|
Search Alerts/Recalls
|
|