RESMED LTD ASTRAL 100 - AMER Back to Search Results
Model Number 27001
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/12/2022
Event Type  malfunction  
Manufacturer Narrative
The astral device was returned to resmed.Evaluation confirmed the reported complaint.Visual inspection revealed dust and dirt within the device.The pneumatic block was replaced to address the reported complaint.The device was serviced and fully tested before it was returned to the customer.Resmed reference#: (b)(4).
 
Event Description
It was reported to resmed that an astral device displayed an error message (sf147) related to the main blower.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 100 - AMER
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
MDR Report Key15569524
MDR Text Key301562940
Report Number3007573469-2022-00669
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00619498270019
UDI-Public(01)00619498270019(11)170606(10)1243046
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27001
Device Catalogue Number27001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA10/10/2022
Distributor Facility Aware Date09/12/2022
Device Age64 MO
Date Report to Manufacturer10/10/2022
Date Manufacturer Received09/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/09/2017
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-