It was reported that during a rotator cuff repair procedure, the suture was being passed through the shoulder with a firstpass device, but the jaw locked closed inside the patient.In consequence, the surgeon ended up removing the device by tearing through the patient rotator cuff tissue.The procedure was successfully completed using a competitor device.There was a non-significant surgical delay of 10-15min, and no further complications were reported.
|
H10: internal complaint reference: (b)(4).H3, h6: the reported device was received for evaluation.A visual evaluation showed the device was not returned with any original packaging.The lever and trigger are fully compressed against the handle.The suture passer needle is fully extended and pulled outside of the jaw.The jaw is locked down over the needle.The suture capture has no defect.Bio debris is present.A functional evaluation revealed the lever and trigger will not release.Nor will the jaw open with manual pressure.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with a mechanical component failure.Factors that could have contributed to the failure include rough handling or excessive force to the device.No containment or corrective actions are recommended at this time.
|