AESCULAP AG CHALLENGER TI-P SM-LIGAT.CLIPS 12 CARTR.; LIGATION / VESSEL CLIPS Back to Search Results
Model Number PL574T
Device Problems Fracture (1260); Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/28/2022
Event Type  malfunction  
Event Description
It was reported that there was an issue with pl574t - challenger ti-p sm-ligat.Clips 12 cartr.According to the complainant, after the cartridge was used, it was taken out and the proximal end was found broken.The fragment was suspected to have remained in situ, in the patient's abdomen.The adverse event / malfunction is filed under aag reference (b)(4).
 
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
 
Manufacturer Narrative
Additional information: b5 - event clarification.Investigation results: visual investigation: here we detected a broken off latch, a deformed tip of the slider sheet, and wrong positioned clips.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.There are no similar complaints against the same lot number(s) with this error pattern.The review of risk assessment revealed that the overall risk level (severity 3(5) x probability of occurrence 1(5)) according to din en iso 14971 is still acceptable.Explanation and rationale: according to the quality standard and dhr files a material defect and a production error can be excluded.There are no hints of a pre- damage or something similar.Investigation leads to the assumption that the cause for the mentioned deviation was caused by a improper removal of the titanium clip cartridge from the challenger shaft.If the magazine was removed not according to the instructions for use (ifu), there is the possibility that the latch broke off.In addition, a rough removal of the magazine from the sterile packaging could also lead to a broken off latch.Furthermore, the deformed tip of the slider sheet was most likely caused by the jaws being closed while the slider sheet was still in the jaws to forward a clip.Furthermore, according to the ifu, certain points must be observed such as details on inserting and removing the titanium clip cassette.Conclusion and measures / preventive measures: based upon the investigation results, the root cause of the problem is most probably usage-related.Based upon the investigations results a capa is not necessary.
 
Event Description
Update: the piece was not recovered.The patient had no negative effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHALLENGER TI-P SM-LIGAT.CLIPS 12 CARTR.
Type of Device
LIGATION / VESSEL CLIPS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key15579814
MDR Text Key302261085
Report Number9610612-2022-00299
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
K081031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPL574T
Device Catalogue NumberPL574T
Device Lot Number52700958
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/24/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-