• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH AO/ASIF© REAMING ATTACHMENT F/BPL; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH AO/ASIF© REAMING ATTACHMENT F/BPL; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.780
Device Problems Component Missing (2306); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2022
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The device manufacture date is unknown.Additional narrative: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to improper maintenance.Udi: (b)(4).
 
Event Description
It was reported that during service and evaluation, it was determined that the attachment device came apart - unattached, had a corroded bearing and missing components.It was further determined that the device failed pretest for general condition, check function of the tool coupling and check tool coupling.It was noted in the service order that the device was undone, missing a spring.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2022.All available information has been disclosed.If additional information should become available, an additional report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AO/ASIF© REAMING ATTACHMENT F/BPL
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key15581361
MDR Text Key302598161
Report Number8030965-2022-07182
Device Sequence Number1
Product Code GEY
UDI-Device Identifier07611819137219
UDI-Public07611819137219
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.780
Device Lot Number3548470
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2022
Date Manufacturer Received09/13/2022
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-