• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MAESTRO 3000 CARDIAC ABLATION SYSTEM - CONTROLLER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MAESTRO 3000 CARDIAC ABLATION SYSTEM - CONTROLLER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Model Number 86245
Device Problem Output below Specifications (3004)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/01/2022
Event Type  malfunction  
Manufacturer Narrative
The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported a maestro 3000 cardiac ablation system was selected for use.Equipment presents a lot of interference in the intracavitary signals of the therapeutic catheter, even placing grounding may be a problem of current leakage and this interference worsens with the application of radio frequency (rf) ablation.In addition, the power drops when rf is applied, the power does not increase and not able to reach the programmed power output.It is unknown if this led to incomplete or inadequate ablation.No further patient or procedure information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAESTRO 3000 CARDIAC ABLATION SYSTEM - CONTROLLER
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
STELLARTECH RESEARCH CORPORATION
560 cottonwood drive
milpitas CA 95035
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key15590381
MDR Text Key302067547
Report Number2124215-2022-41144
Device Sequence Number1
Product Code OAD
Combination Product (y/n)N
Reporter Country CodeBR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative,Distributor
Reporter Occupation 505
Type of Report Initial
Report Date 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number86245
Device Catalogue Number86245
Device Lot Number021313004M
Was Device Available for Evaluation? No
Date Manufacturer Received09/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/13/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-