Model Number HL-90
Device Problems Device Sensing Problem (2917); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/29/2022
Event Type  malfunction  
Manufacturer Narrative
Udi and operator of device are unknown.No information has been provided to date.A product sample was received and is awaiting evaluation and investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the water level sensor is not working.There has been no report of patient involvement or no observable clinical symptoms or a change in symptoms identified in the patient.
 
Manufacturer Narrative
Other text: the device was returned for evaluation.The physical condition of the device indicated wear.Specifically, the front cover had a large chip in it, a discolored line cord, a corroded quick connect, and scratches on the water tank.Functional testing was performed; the float switch performed correctly but found the water tank cover cracked.The complaint was not confirmed.Repairs and/or replacements were completed.Preventive maintenance was performed.A manufacturing device history record review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.(h6c, h6d, h6e).
 
Manufacturer Narrative
Other text: additional event information received indicating no patient involvement and date of event was received.H6: event problem and evaluation codes: updates not required., corrected data: h6: correction: event problem device code:a0709 [2917], device sensing problem c63238.
 
Event Description
Additional info received via email 17-oct-2022: event date updated: 08/29/2022.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE LOW FLOW SYSTEMS - HL-90
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15591625
MDR Text Key306685599
Report Number3012307300-2022-23606
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/04/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/24/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-