Model Number HL-90 |
Device Problems
Device Sensing Problem (2917); Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/29/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Udi and operator of device are unknown.No information has been provided to date.A product sample was received and is awaiting evaluation and investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that the water level sensor is not working.There has been no report of patient involvement or no observable clinical symptoms or a change in symptoms identified in the patient.
|
|
Manufacturer Narrative
|
Other text: the device was returned for evaluation.The physical condition of the device indicated wear.Specifically, the front cover had a large chip in it, a discolored line cord, a corroded quick connect, and scratches on the water tank.Functional testing was performed; the float switch performed correctly but found the water tank cover cracked.The complaint was not confirmed.Repairs and/or replacements were completed.Preventive maintenance was performed.A manufacturing device history record review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.(h6c, h6d, h6e).
|
|
Manufacturer Narrative
|
Other text: additional event information received indicating no patient involvement and date of event was received.H6: event problem and evaluation codes: updates not required., corrected data: h6: correction: event problem device code:a0709 [2917], device sensing problem c63238.
|
|
Event Description
|
Additional info received via email 17-oct-2022: event date updated: 08/29/2022.No patient involvement.
|
|