The reported event could be confirmed, since the device was returned and is indeed broken.The device inspection could confirm the catalog and lot numbers reported were correct, and that tip of the broach is indeed broken.The device shows various scratches, which could be attributed to multiple years of use and reprocessing.The tip of the device shows a brittle fracture, showing that the instrument broke in a sudden manner, following the application of an important force.Based on investigation, the root cause was attributed to a user related issue.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
|