• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. MAINTENANCE UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. MAINTENANCE UNIT Back to Search Results
Model Number MU-1
Device Problems Break (1069); Material Protrusion/Extrusion (2979)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/13/2022
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer¿s allegation was confirmed.During the inspection it was found that the power switch and cover was broken exposing the internal metal components.In addition, it was found that the power switch was also corroded, the suction feet was worn and lost, the feet screws and rear panel were contaminated, the output socket was dirty, the chassis was rusty and contaminated, and the top cover paint was peeling off.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if additional information becomes available.
 
Event Description
The customer reported to olympus that the green button on front of the maintenance unit was found to be broken during maintenance.There were no reports of patient harm.During inspection and testing of the returned unit, there was a broken on/ off switch in which the internal metal components were exposed.This mdr is being submitted to capture the reportable malfunction found in the device evaluation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the following led to the metallic part of the internal circuit being exposed and the power switch and cover breaking: a) excessive force.B) repeated continued use.A definitive root cause cannot be identified.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAINTENANCE UNIT
Type of Device
MAINTENANCE UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer Contact
masaharu hirose
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8-061
JA   961-8061
426422891
MDR Report Key15596772
MDR Text Key307081282
Report Number3002808148-2022-03121
Device Sequence Number1
Product Code FDF
UDI-Device Identifier04953170060434
UDI-Public04953170060434
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K051645
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMU-1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/01/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-