Device analysis: the returned product consisted of a 35mm watchman flx delivery system (wds) with the implant deployed.The implant, core wire, tip, sheath, and hub were microscopically and visually analyzed.Multiple kinks were noted in the sheath.Inspection of the tip found flared tri-cuts and abrasions on the inside of the sheath tip.The closure device anchors were hooked onto the struts.Product analysis could not confirm the reported events of difficulty recapturing and deploying as clinical circumstances could not be replicated.Product analysis confirmed the reported event of struts twisted, as the closure device anchors were hooked onto the struts.
|