Model Number CLV-190
Device Problem Moisture or Humidity Problem (2986)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Before a procedure, the customer reported intermittent errors b30 and e216 (scope communication error) on several endoscopes.Occasionally, the endoscope would give the error, but it would work after being disconnected and reconnected.The evis exera iii xenon light source was very humid inside and its screws were broken.The connector cannot be removed.There were no reports of patient harm associated with this event.
 
Manufacturer Narrative
The device has not been received to date.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 9 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a definitive root cause of the b30 error could not be identified.However, it's likely the cause is related to rust in the front connection and inside the unit.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer Contact
masaharu hirose
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8-061
JA   961-8061
426422891
MDR Report Key15611587
MDR Text Key307162402
Report Number3002808148-2022-03197
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/07/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-