|
Model Number AIA-2000 ST |
Device Problem
Priming Problem (4040)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/26/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
A field service engineer (fse) was at the customer¿s site to address the reported event.Fse confirmed the complaint by reviewing the error logs and reproduced the error by running the daily check and the wash probes did not completely remove the wash from the well wash.Fse resolved the complaint by replacing the bf waste pump 2.Fse repaired and validated the analyzer by successfully performing quality control run, which passed within acceptable range.No further action required by field service.The aia-2000 analyzer is functioning as expected.The aia-2000, serial number (b)(4), was installed on (b)(6) 2022.A complaint and service history review for similar complaints was performed from installation date (b)(6) 2022through aware date (b)(6) 2022.There were no other similar complaints identified during the searched period.Aia-2000 operator's manual on the appendix 4: error messages: (2231) b/f probe 1 overflow sensor failure cause: the overflow sensor is detecting liquid continuously.Solution: clean b/f probe 1.(see chapter 9, "maintenance") if retry fails, contact tosoh service center or local representatives.(2233) b/f probe 3 overflow sensor failure cause: the overflow sensor is detecting liquid continuously.Solution: clean b/f probe 3.(see chapter 9, "maintenance") if retry fails, contact tosoh service center or local representatives.(2232) b/f probe 2 overflow sensor failure cause: the overflow sensor is detecting liquid continuously.Solution: clean b/f probe 2.(see chapter 9, "maintenance") if retry fails, contact tosoh service center or local representatives.The most probable cause of the reported event was due to the faulty bf waste pump 2.
|
|
Event Description
|
A customer reported error messages 2231 b/f probe 1 overflow sensor failure, 2232 b/f probe 2 overflow sensor failure, and 2233 b/f probe 3 overflow sensor failure on the aia-2000 analyzer.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event which resulted in a delayed reporting of patient samples for alpha-fetoprotein (afp).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|
|
Manufacturer Narrative
|
A review of the device history record (dhr) was conducted for serial number 10651901, which confirmed that there were no nonconformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported bf waste pump 2 failure.
|
|
Manufacturer Narrative
|
The drain pump was returned to tosoh instrument service center for investigation.Functional testing confirmed the reported event was due to failure of the waste pump.
|
|
Search Alerts/Recalls
|
|
|