Model Number 3300
Device Problems Display or Visual Feedback Problem (1184); Device Sensing Problem (2917); Key or Button Unresponsive/not Working (4063)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/29/2019
Event Type  malfunction  
Event Description
It was reported that during an implant procedure, this programmer exhibited noise while using the pacing system analyzer (psa).The cables were removed and placed into a different psa with no further noise observed.During a different procedure the following week it was noted that the programmer screen was frozen.For threshold testing the voltage was set to 5v, however an impedance measurement was unable to be obtained.The threshold was able to decrement however on the screen it still showed 5v although loss of capture was observed on a surface.A different model programmer and psa were used.No adverse patient effects were reported.
 
Manufacturer Narrative
The device was not returned for product analysis as the programmer remains in service at the healthcare facility.Although the device was not returned for analysis, boston scientific performed an investigation with all available information.An unresponsive touchscreen is a known and anticipated potential hazard that can occur during use of the device which is accounted for in device risk documentation and labeling.Boston scientific has received similar reported events where the latitude programmer screen became unresponsive to touch and a complaint review confirmed that the occurrence of harm associated with these events is low.Investigation of this behavior has not found any commonalities (i.E., device serial numbers, date of the event, date of manufacture, etc.) in the reports received.
 
Event Description
It was reported that during an implant procedure, this programmer exhibited noise while using the pacing system analyzer (psa).The cables were removed and placed into a different psa with no further noise observed.During a different procedure the following week it was noted that the programmer screen was frozen.For threshold testing the voltage was set to 5v, however an impedance measurement was unable to be obtained.The threshold was able to decrement however on the screen it still showed 5v although loss of capture was observed on a surface.A different model programmer and psa were used.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LATITUDE PROGRAMMING SYSTEM
Type of Device
EXTERNAL PROGRAMMER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
PLEXUS CORPORATION
2400 millbrook dr
buffalo grove IL 60089
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15615557
MDR Text Key306358955
Report Number2124215-2022-41725
Device Sequence Number1
Product Code OSR
UDI-Device Identifier00802526608513
UDI-Public00802526608513
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910077/S159
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 06/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/17/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3300
Device Catalogue Number3300
Device Lot Number007814
Was Device Available for Evaluation? No
Date Manufacturer Received05/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-