It was reported that during use the intra-aortic balloon pump (iabp) suddenly showed alarm of optic sensor failure.Instructions to trouble shoot the device were followed, but the alarm continued.The customer spoke with device representative but the issue was not resolved.Throughout the rest of the shift, every pvc/pac beat the patient had on the bedside monitor, the unit was unable to receive an appropriate trigger and the device would automatically default into "standby" mode, thus not supporting the patient.The patient was experiencing more ectopy thereafter and the unit had to be manually restarted every few seconds.During this time, the patient denied any pain/discomfort.The patient was taken to the cath lab for replacement of the iabp catheter and the console was also changed out.Medwatch # mw5112091.
|
It was later informed by the customer that the unit was placed back into service after evaluation and maintenance which was performed by the customer.The customer also stated that the issue was more about the optical sensor of the catheter rather than the console, even though the console was changed out.No other information was provided regarding the repair.This complaint will be closed, if further information is received in regards to the repair of the unit, the complaint will be opened and updated accordingly.
|