MAUDE Adverse Event Report: SOCLEAN, INC. SOCLEAN 2; SLEEP EQUIPMENT MAINTENANCE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOCLEAN, INC. SOCLEAN 2; SLEEP EQUIPMENT MAINTENANCE SYSTEM Back to Search Results
Model Number SC1200
Device Problems Decrease in Pressure (1490); Appropriate Term/Code Not Available (3191)
Patient Problem Respiratory Tract Infection (2420)
Event Date 09/23/2022
Event Type  Injury  
Manufacturer Narrative
File a 30-day mdr.An assessment was conducted.The event is still considered reportable under fda's regulation.A clinical investigation was attempted.The customer was unwilling to give other important information related to the injury, making it difficult to determine the cause of the respiratory infection.Reporting for due diligence.
 
Event Description
Customer reports several respiratory infections.Md seen and antibiotic prescribed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOCLEAN 2
Type of Device
SLEEP EQUIPMENT MAINTENANCE SYSTEM
Manufacturer (Section D)
SOCLEAN, INC.
1 vose farm road
peterborough NH 03458
Manufacturer (Section G)
SOCLEAN, INC.
1 vose farm road
peterborough NH 03458
Manufacturer Contact
jessica wilson
1 vose farm road
peterborough,, NH 03458
6033712570
MDR Report Key15624115
MDR Text Key301906981
Report Number3009534409-2022-00585
Device Sequence Number1
Product Code LRJ
UDI-Device Identifier00187293000860
UDI-Public187293000860
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberSC1200
Device Catalogue NumberSC1200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2022
Date Manufacturer Received09/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
-
-