• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION TRAPEZOID RX; LITHOTRIPTOR, BILIARY MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION TRAPEZOID RX; LITHOTRIPTOR, BILIARY MECHANICAL Back to Search Results
Model Number M00510880
Device Problems Break (1069); Difficult to Open or Close (2921); Material Deformation (2976); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/01/2022
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that a trapezoid rx basket was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on an unknown procedure.During the procedure, the basket failed to close.The procedure was completed with another trapezoid rx.There were no patient complications as a result of this event.Investigation results revealed the handle cannula detached/separated; therefore, this is now an mdr reportable event.
 
Manufacturer Narrative
The exact date of the event is unknown.The provided event date was chosen as a best estimate based on the date that the manufacturer became aware of the event.The complainant was unable to provide the suspect device lot number; therefore, the expiration and device manufacture dates are unknown.(b)(4).The returned trapezoid basket was analyzed, and a visual evaluation noted that the handle cannula was detached and had drag marks of the fastening screws.This condition does not allow the basket to open or close.Additionally, the side car rx was pushed back.The depth of the fastening screws was measured and the depth of the "proximal screw" was outside of the allowed tolerance.The reported event was confirmed.Based on all available information, the most likely cause is attributable to a manufacturing deficiency.Due that the proximal screw does not have the suitable depth, (the depth was measured at 4 different points and all of them were out of specification), the proper fastening was not given.The force applied when activating the device causes the detachment of the handle cannula leaving drag marks of the screws on it.The detachment of the handle cannula in conjunction with the manipulation during the procedure could have caused adverse events as consequence, resulting in the failure to close the basket and the side car rx push back.Therefore, the most probable root cause is adverse event related to procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEZOID RX
Type of Device
LITHOTRIPTOR, BILIARY MECHANICAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15625670
MDR Text Key302186856
Report Number3005099803-2022-06121
Device Sequence Number1
Product Code LQC
UDI-Device Identifier08714729296393
UDI-Public08714729296393
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K040447
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00510880
Device Catalogue Number1088
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/23/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-