• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH GSTRO FEED TBE W/Y PRT 20FR EN; GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH GSTRO FEED TBE W/Y PRT 20FR EN; GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS Back to Search Results
Model Number 8884720205E
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/10/2022
Event Type  malfunction  
Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event Description
The customer reported that they have been using the kangaroo balloon gastrostomy tubes successfully in a patient for several months now, however he has started to have issues with carers breaking the adapter at the end of the tube.No injuries were reported.Per additional information provided on october 17, 2022, the y-port cracked and ended up leaking becoming unusable.The tube was unusable until the adapter was changed.The adapter was removed from another tube and replaced onto the tube in situ as the patient is unable to easily have his tube changed and it should be avoided when possible.This occurred with two of three tubes.Further information provided on october 18, 2022 stated that when the initial y-port cracked and leaked, the port was removed off of the tube and replaced with another y-port that then cracked and leaked as well.The second y-port was then removed and replaced with another y-port from a third tube.
 
Manufacturer Narrative
A device history record review could not be performed because a lot number was not received with the complaint.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to the release of product.A physical sample was not received for the investigation, but a photo was provided.The photo was visually inspected, but it was not possible to confirm the reported condition through the photo.However, a corrective and preventative action has been initiated to address the reported condition through a more robust investigation.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GSTRO FEED TBE W/Y PRT 20FR EN
Type of Device
GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key15626113
MDR Text Key303404156
Report Number9612030-2022-03427
Device Sequence Number1
Product Code PIF
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8884720205E
Device Catalogue Number8884720205E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received10/10/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage A
Patient Sequence Number1
Patient SexMale
-
-