Model Number EPIQ 7 DIAGNOSTIC ULTRASOUND SYSTEM |
Device Problem
No Device Output (1435)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/28/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
An investigation is underway but has not been completed.The results of the investigation will be included in the final report.
|
|
Event Description
|
A customer reported their epiq 7c ultrasound system became unresponsive with an 007 error code displayed during an unspecified procedure while using an x8-2t transducer.The procedure in progress was completed successfully after disconnecting and reconnecting the transducer multiple times to clear the code.There was no patient or user harm associated with this event.
|
|
Manufacturer Narrative
|
The field service engineer (fse) replaced the x8-2t transducer.A thorough technical investigation of the suspected transducer was performed.The engineering team found the transducer with cosmetic damage.The midhandle cosmetic specs failed, the sensor cosmetic specs failed due to scratches on the tip / cap.The transducer passed the image, electrical and articulation tests.There was no reproducible error code generated.
|
|