• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS ULTRASOUND EPIQ 7; SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS ULTRASOUND EPIQ 7; SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC Back to Search Results
Model Number EPIQ 7 DIAGNOSTIC ULTRASOUND SYSTEM
Device Problem No Device Output (1435)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/28/2022
Event Type  malfunction  
Manufacturer Narrative
An investigation is underway but has not been completed.The results of the investigation will be included in the final report.
 
Event Description
A customer reported their epiq 7c ultrasound system became unresponsive with an 007 error code displayed during an unspecified procedure while using an x8-2t transducer.The procedure in progress was completed successfully after disconnecting and reconnecting the transducer multiple times to clear the code.There was no patient or user harm associated with this event.
 
Manufacturer Narrative
The field service engineer (fse) replaced the x8-2t transducer.A thorough technical investigation of the suspected transducer was performed.The engineering team found the transducer with cosmetic damage.The midhandle cosmetic specs failed, the sensor cosmetic specs failed due to scratches on the tip / cap.The transducer passed the image, electrical and articulation tests.There was no reproducible error code generated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIQ 7
Type of Device
SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC
Manufacturer (Section D)
PHILIPS ULTRASOUND
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS ULTRASOUND
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
eric nail
22100 bothell everett highway
bothell, WA 98021
MDR Report Key15626701
MDR Text Key303559263
Report Number3019216-2022-00093
Device Sequence Number1
Product Code IYN
UDI-Device Identifier00884838097872
UDI-Public00884838097872
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132304
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEPIQ 7 DIAGNOSTIC ULTRASOUND SYSTEM
Device Catalogue Number795201
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/30/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/14/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-