The reporter contacted abbott diabetes care, alleging the meter will not turn on.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no report of death, serious injury, or mistreatment associated with this event.
|
To date, product has not been returned for further investigation.The useful life of freestyle mini meter is 5 years.As the manufacturing date of this product is 2006, it has been in distribution beyond its useful life at the time of the complaint.Since the product exceeded its useful life, it is determined to have met specification when the product was released and through its lifespan.No further investigation activities are being performed at this time.All pertinent information available to abbott diabetes care has been submitted.
|