This report has been identified as b.Braun medical internal report number (b)(4).One (1) used sample with the packaging was returned for evaluation.The sample was subject to a visual and physical evaluation with passing results.No leakage was observed on the discofix.Based on the evaluation results, the reported defect was not confirmed.Retained units were evaluated and passed the internal testing.Review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|