One hemosphere instrument was returned csm 16802554, pc2k 600101035534, hrs 600104067265, a power cord, and a cuff.These were all connected to the hem1 except the cuff because it is expired.The connected hrs successfully zeroed and was able to acquire a normal waveform and bp readings on both cuff ports using a finger/cuff simulator.The hem1 monitor connected to and recognized a swan ganz module on module ports 1 and 2 and was also able to connect to and recognize an oximetry cable on smart cable ports 1 and 2.No error messages were observed.No damage was found.The hem1 and hemcsm10 passed electrical safety test.A visual inspection was done on the cuff and a hole was found in the plastic that covers the electrical components.This is an observation only and it can't be confirmed to be related to the rfr.
|