C.R. BARD, INC. (BASD) -3006260740 SITE RITE 8 ULTRASOUND SYSTEM 20MM CUE LINEAR PROBE; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770603
Device Problems No Display/Image (1183); Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/13/2022
Event Type  malfunction  
Event Description
Half of the image is not showing up on the screen.
 
Manufacturer Narrative
The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The device has not been received by the manufacturer for evaluation.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of half of the image is not showing up on the screen was confirmed.The image would show missing segments.The root cause of the reported issue is a probe failure.
 
Event Description
Half of the image is not showing up on the screen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE RITE 8 ULTRASOUND SYSTEM 20MM CUE LINEAR PROBE
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key15634818
MDR Text Key306998435
Report Number3006260740-2022-04079
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741138218
UDI-Public(01)00801741138218
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182281
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770603
Device Catalogue Number9770603
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/10/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-