MAUDE Adverse Event Report: SIRONA DENTAL SYSTEMS GMBH MIDWEST E PLUS 1:5 HIGH SPEED CONTRA ANGLE ATTACHMENT; HANDPIECE, AIR-POWERED, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIRONA DENTAL SYSTEMS GMBH MIDWEST E PLUS 1:5 HIGH SPEED CONTRA ANGLE ATTACHMENT; HANDPIECE, AIR-POWERED, DENTAL Back to Search Results
Catalog Number 875205
Device Problem Overheating of Device (1437)
Patient Problems Burn(s) (1757); Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Additional serial numbers included in this report: (b)(4).27 of 27 devices were returned for evaluation.Evaluation of 5 devices found them all to be within specification.Evaluation of one device found excessive wear due to a lack of proper maintenance.The device did not heat up during evaluation.The device was repaired and returned to the customer.Evaluation of one device found excessive wear due to a lack of proper maintenance.The device had debris build-up.The device did not heat up during evaluation.The device was repaired and returned to the customer.Evaluation of four devices found excessive wear due to a lack of proper maintenance.A friction problem was identified from pressure on the cap for these devices.The devices did not heat up during evaluation.The devices were repaired and returned to the customers.Evaluation of four devices found excessive wear due to a lack of proper maintenance.A friction problem was identified from pressure on the cap for these devices.The devices did heat up during evaluation.The devices were repaired and returned to the customers.Evaluation of six devices found excessive wear due to a lack of proper maintenance.A friction problem was identified from pressure on the cap for these devices.The devices had debris build-up.The devices did not heat up during evaluation.The devices were repaired and returned to the customers.Evaluation of one device found excessive wear due to a lack of proper maintenance and lubrication.A friction problem was identified from pressure on the cap.The device did heat up during evaluation.The device was repaired and returned to the customer.Evaluation of four devices found excessive wear due to a lack of proper maintenance and lubrication.A friction problem was identified from pressure on the cap.The devices had debris build-up.The devices did not heat up during evaluation.The devices were repaired and returned to the customers.Evaluation of one device found excessive wear due to a lack of proper maintenance.A friction problem was identified from pressure on the cap.This device had a deformation issue (dent).The device had debris build-up.The device did not heat up during evaluation.The device was repaired and returned to the customer.
 
Event Description
This report summarizes 27 malfunction events where a midwest e plus 1:5 high speed contra angle attachment overheated.Three patients experienced minor burns, no intervention was required.There was no injury in 23 of the events.There was one event where the patient outcome is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDWEST E PLUS 1:5 HIGH SPEED CONTRA ANGLE ATTACHMENT
Type of Device
HANDPIECE, AIR-POWERED, DENTAL
Manufacturer (Section D)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM  D-64625
Manufacturer (Section G)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM   D-64625
Manufacturer Contact
hannah seevaratnam
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494593
MDR Report Key15634891
MDR Text Key304716919
Report Number9614977-2022-00062
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
PMA/PMN Number
K972436
Number of Events Reported27
Summary Report (Y/N)Y
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 10/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number875205
Patient Sequence Number1
-
-