SMITH & NEPHEW, INC. LGN HOUSING RESECT COLLET S3-8; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 71434424 |
Device Problems
Fracture (1260); Separation Failure (2547)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/27/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that, during a tka surgery, it was noticed that an lgn housing resect collet s3-8 had become stuck on the box chisel instrument, due to it splintering.The procedure was resumed, without any delay, changing the surgical technique.No other complications were reported.
|
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).
|
|
Manufacturer Narrative
|
Section h10: the device was not returned for evaluation and the reported event could not be confirmed.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Internal complaint reference number: (b)(4).
|
|
Event Description
|
It was reported that, during a tka surgery, it was noticed that an lgn housing resect collet s3-8 had become stuck on the box chisel instrument, due to it splintering.The procedure was resumed, without any delay, by using a double action nibbler.Nothing fell in the patient.No other complications were reported.
|
|
Event Description
|
It was reported that, during a tka surgery, it was noticed that an lgn housing resect collet s3-8 had become stuck on the box chisel instrument, due to it splintering.The procedure was resumed, without any delay, by using a double action nibbler.Nothing fell in the patient.No other complications were reported.
|
|
Manufacturer Narrative
|
Additional information: b5.
|
|