• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OAKDALE LEVEL 1 FAST FLOW FLUID WARMER; WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 FAST FLOW FLUID WARMER; WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION Back to Search Results
Catalog Number H-1000-INT-100
Device Problem Device Damaged Prior to Use (2284)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/06/2018
Event Type  malfunction  
Event Description
It was reported that when assembling the product, the air hose connection port's plug got damaged.No patient injury was reported.
 
Manufacturer Narrative
This mdr was generated under protocol (b)(4) as a result of warning letter cms# (b)(4).Manufacturing device history record review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.Visual and functional testing were performed.Visual inspection observed that the top and bottom cases were in excellent condition.There was no visible contamination.During power up testing (using biomed diagnostics to check digital sensors), the reported problem was duplicated.The root cause of the reported issue was found to be the q1 had broken off from the plunger board.Plunger board also came off the glue.Replaced plunger board.Performed power up process, calibrated device and device woks normally.Udi are unknown.No product information has been provided to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 FAST FLOW FLUID WARMER
Type of Device
WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15637910
MDR Text Key303232276
Report Number3012307300-2022-25333
Device Sequence Number1
Product Code BSB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
BK940056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberH-1000-INT-100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2018
Was the Report Sent to FDA? No
Date Manufacturer Received09/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/09/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-