Catalog Number C37091718-NL
Device Problem Material Perforation (2205)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
(udi) and (510k) are unknown; no additional information was provided.This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.One (1) sample was received without its original packaging.The sample was visually inspected at a distance of 12 inches under normal lighting to received unit, in order to detect any damage on the unit; it was found that there were holes in bag.The most probable root causes are: bag was sent by supplier in defective conditions and production personnel did not detect this issue during the assembly process; or bag was damaged during the manufacturing process.Action(s) taken to mitigate the reported issue(s): as preventive action, production personnel was notified by area trainer as awareness of the defect and to ensure that operators verify that breathing bags are free of holes and damage.
 
Event Description
It was reported that multiple breathing bags in the custom circuits have small tears or perforations in the non-latex 3l bag.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15638648
MDR Text Key306923397
Report Number3012307300-2022-25371
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/17/2021
Device Catalogue NumberC37091718-NL
Device Lot Number3565034
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2018
Was the Report Sent to FDA? No
Date Manufacturer Received10/03/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/19/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-