(udi) and (510k) are unknown; no additional information was provided.This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.One (1) sample was received without its original packaging.The sample was visually inspected at a distance of 12 inches under normal lighting to received unit, in order to detect any damage on the unit; it was found that there were holes in bag.The most probable root causes are: bag was sent by supplier in defective conditions and production personnel did not detect this issue during the assembly process; or bag was damaged during the manufacturing process.Action(s) taken to mitigate the reported issue(s): as preventive action, production personnel was notified by area trainer as awareness of the defect and to ensure that operators verify that breathing bags are free of holes and damage.
|