Device Problems Therapeutic or Diagnostic Output Failure (3023); Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/13/2022
Event Type  malfunction  
Event Description
It was reported that, the cardiosave intra-aortic balloon pump (iabp) unit had an error code 4561.There was no patient involvement.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Manufacturer Narrative
Updated field: b4, d9, e1-(site country, site email), g3, g6, h2, h3, h4, h6(medical device ¿ problem code, type of investigation, investigation findings, investigation conclusions),h10.It was reported that the cs300 intra-aortic balloon pump (iabp) unit had error code 4561.Customer complained that during patient transport the unit shut down.No patient death or injury reported.A getinge field service engineer (fse) evaluated the unit and found the k3/k5 assy was loose and touched the solenoid driver board causing it to short.There was no damage to any parts.Fse secured the assy and also performed a pm during this call.Unit passed all functional and safety tests per factory specifications.Returned unit to the customer and cleared for clinical use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key15640123
MDR Text Key304087250
Report Number2249723-2022-02723
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received03/18/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/26/2001
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-