Catalog Number 8606500 |
Device Problems
Decrease in Pressure (1490); Inaccurate Delivery (2339); No Pressure (2994)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/04/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that duriing a case the airway pressure was lost.There was no injury reported.The device reportedly generated an alarm.
|
|
Manufacturer Narrative
|
The investigation has just started; results will be provided in a follow-up report.
|
|
Event Description
|
It was reported that duriing a case the airway pressure was lost.There was no injury reported.The device reportedly generated an alarm.
|
|
Manufacturer Narrative
|
The electronic log file was available for investigation.The case in question could be reconstructed by means of the information stored therein.Beginning at 5:07pm minute volume leaks of up to 4.4 l/min in the patient circuit were present.The log entries do not indicate a device failure.The device detected the external leak and reacted as specified for such situation as it posted several alarms (e.G.Fresh gas low or leakage, apnea).
|
|