MEDTRONIC MED REL MEDTRONIC PUERTO RICO ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
|
Back to Search Results |
|
Model Number 37612 |
Device Problems
Break (1069); Delayed Charge Time (2586); Impedance Problem (2950); Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/01/2022 |
Event Type
Injury
|
Event Description
|
It was reported that the patient had the implantable neurostimulator (ins) implanted in a different country in (b)(6) 2010.In august 2022, it was reported that since july, the ins was taking longer to recharge than normal.No falls were reported.Impedances were out of range.When the battery was explanted, it was noted that there was a dent on the surface of the battery.The patient is fine after battery replacement and all impedances are normal.
|
|
Manufacturer Narrative
|
The event date is an estimated date.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3: analysis of the ins showed that it was functioning within specification but had reduced capacity due to overdischarge.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|