• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Catalog Number CON-HL-90
Device Problem Device Damaged Prior to Use (2284)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).Manufacturing device history record review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.Visual inspection observed that the device was in poor condition with external labels and components.During functional testing, the reported problem was duplicated.The root cause of the reported issue was found to be the normal wear and tear on the power switch.Due to the age and upgrades needed, the device was deemed beyond economical repair and will be scrapped.A corrective and preventative action has been opened to address the reported issue.Udi) is unknown.No product information has been provided to date.
 
Event Description
It was reported that the damage power switch.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
2823 franklin rd sw bldg a
minneapolis, MN 55442
MDR Report Key15651765
MDR Text Key306893285
Report Number3012307300-2022-25737
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/23/2022
Initial Date FDA Received10/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/28/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-