Model Number 4543
Device Problems Failure to Capture (1081); Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/14/2022
Event Type  Injury  
Manufacturer Narrative
Information indicates this lead will be returned.This investigation will be updated should further information be provided.
 
Event Description
It was reported that this left ventricular (lv) lead exhibited failure to capture at maximum pacing outputs.An x-ray was performed, and an obvious kink in the lead body was observed, indicating the lead is fractured.As a result, the lv lead was explanted and replaced.There were no additional adverse patient effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2 IS-1
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15654175
MDR Text Key302245024
Report Number2124215-2022-42939
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526410994
UDI-Public00802526410994
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P010012/S024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/25/2014
Device Model Number4543
Device Catalogue Number4543
Device Lot Number175643
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/25/2012
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age58 YR
Patient SexMale
-
-