Model Number 7400S |
Device Problem
Break (1069)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 09/23/2022 |
Event Type
malfunction
|
Event Description
|
Medtronic received information that when the sizer was placed in position, it broke away from the handle introducer into the patient.It was stated that the broken off sizer was retrieved from the patient by hand.It was stated that a different product was used to complete the procedure.It was mentioned the sizers have been used since the inception of the avalus valve at the facility.The cleaning process consists of cleaning, disinfecting in a washer, visual check and steam sterilization.The products used during this process include medirinse rinse aid and genesis detergent.No additional adverse patient effects were reported and the patient has since been discharged.
|
|
Manufacturer Narrative
|
The product has been returned for analysis.A supplemental report will be filed upon completion of the analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Product analysis: upon receipt at medtronic's quality laboratory, visual examination revealed that the barrel end was broken.The replica end exhibited signs of crazing, which is consistent with historical events associated with multiple sterilization cycles.Signs of crazing were observed on the barrel end.Conclusions: the investigation is still in progress.A supplemental report will be filed upon its completion.H3: device evaluated? updated h6: code updated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|