MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX DISPOSABLE ANESTHESIA BREATHING CIRCUIT; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX DISPOSABLE ANESTHESIA BREATHING CIRCUIT; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Catalog Number C45041345-NLJ
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/01/2019
Event Type  malfunction  
Manufacturer Narrative
Udi is unknown.No information provided to date.Premarket (510k) number is unknown.This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No lot number was provided; therefore, device history record review could not be completed.A product sample was received for evaluation.Visual and functional testing were performed.The samples were visually inspected at a distance of 12 under normal lighting to received unit, in order to detect any damage on the unit.The breathing bag have a hole.A review of the manufacturing process for was conducted by quality engineer in order to verify that there are no situations or practices that could create the event.No discrepancies were found.A random sample of (b)(4) units was taken from the manufacturing process, in order to perform a visual inspection to verify that all components were free of damage (scuffs, pinch marks, etc.), cracks, crazing, cuts or other workmanship defects that can affect assembly function or appearance.All the inspections and verifications are being performed by the production personnel and verified by the quality inspectors.The root cause of the reported issue was that the production personnel did not detect and scrap the damaged breathing bag.Actions were taken to mitigate the reported issue: a quality alert (qalert) was generated on to ensure that production personnel detect any marks or damage on the breathing bags.Production personnel was trained on the qalert.
 
Event Description
It was reported that during a pre-use check, the customer noticed air was leaking from the anesthesia circuit or bag.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX DISPOSABLE ANESTHESIA BREATHING CIRCUIT
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
2-12-12 kitanagaseomote-machi,
minneapolis, MN 55442
MDR Report Key15656239
MDR Text Key307042568
Report Number3012307300-2022-25886
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC45041345-NLJ
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2019
Was the Report Sent to FDA? No
Date Manufacturer Received09/29/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-