Model Number EC38-I10F2
Device Problems Partial Blockage (1065); Collapse (1099); Failure to Disconnect (2541)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/15/2022
Event Type  malfunction  
Event Description
There is a brush stuck in the op-channel this event occurred at the time of during reprocessing.There was no report of patient harm.
 
Manufacturer Narrative
This device is not distributed in the usa, therefore the pma/510(k) number is not applicable.If additional information becomes available, a supplemental report will be filed with the new information.
 
Manufacturer Narrative
Evaluation summary: based on the provided photograph of the inside of the pipeline of the aircraft, it was determined that the operation channel had buckled and blocked the pipeline, causing the cleaning brush to become stuck during reprocessing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO COLONOCOPE - I10 STANDARD
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key15664790
MDR Text Key306789868
Report Number9610877-2022-00582
Device Sequence Number1
Product Code FDF
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/25/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEC38-I10F2
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/15/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-