Model Number EC38-I10F2 |
Device Problems
Partial Blockage (1065); Collapse (1099); Failure to Disconnect (2541)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/15/2022 |
Event Type
malfunction
|
Event Description
|
There is a brush stuck in the op-channel this event occurred at the time of during reprocessing.There was no report of patient harm.
|
|
Manufacturer Narrative
|
This device is not distributed in the usa, therefore the pma/510(k) number is not applicable.If additional information becomes available, a supplemental report will be filed with the new information.
|
|
Manufacturer Narrative
|
Evaluation summary: based on the provided photograph of the inside of the pipeline of the aircraft, it was determined that the operation channel had buckled and blocked the pipeline, causing the cleaning brush to become stuck during reprocessing.
|
|