Model Number N/A |
Device Problem
Fitting Problem (2183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/29/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that outside of surgery, the blade of the device fell too far back causing the graft thickness adjustment to be inconsistent.There was no patient involvement and as a result no harm or delay were reported.No adverse event was reported as it relates to the event.
|
|
Manufacturer Narrative
|
This incident has been recorded under (b)(4).Upon completion of investigation, a final/supplemental report will be submitted.
|
|
Event Description
|
No additional information is available.
|
|
Manufacturer Narrative
|
This incident has been recorded under (b)(4).Review of the most recent repair record determined the device was out of calibration.The device was recalibrated and the thickness control shaft, sleeve bearings, shaft bearings, and one neck cap screw were replaced to aid in proper calibration and resolved the reported issue.Dhr was reviewed and no discrepancies related to the reported event were found.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|