• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number ASKU
Device Problems No Display/Image (1183); Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/21/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: 2022-60004.
 
Event Description
A customer reported that an ophthalmic operating console stops on the black screen and restarts automatically during setup of a surgery.There was no patient involvement.
 
Manufacturer Narrative
A video of the issue was provided by the customer.The video depicted the boot up issue the customer was experiencing.No additional related information was provided.Specific product identifiers (lot number, batch number, and/or serial number) were not provided and could not be determined at this time.However, all device history records are reviewed prior to product release to ensure the product was manufactured in compliance with the device master record and meets release criteria.A review for complaints reported against this lot/batch/serial number cannot be performed as the lot/batch/serial number is unknown.The lot/batch/serial is unknown; therefore, a service history review cannot be performed.Based on the information obtained, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15671751
MDR Text Key303152794
Report Number2028159-2022-01502
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K120912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 01/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberASKU
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Date Manufacturer Received12/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-